We are China manufacturer of high quality grade 8 Threaded Rods and threaded bars in astm j429 standard. Our advantages and more services of threaded steel rods includes:
1) Wide range of diameters from 1/2" - 4"
2) Customized Length from 100mm to 12000mm
3) Good threads at 60 degree
4) Good straightness: 1mm/m max
5) Even hardness on whole bars
5) Proper mechanical properties
7) Full sets of testing equipment: Surface, Dimenstion. Mechanical properties testing
8) Fast delivery time
Sae J429 Grade 8 Threaded Steel Bar,Grade 8 Threaded Rod,Sae J429 Grade 8 Steel Bar Properties,b7 stud bolt,b7 anchor bolt SHANDONG LE REN SPECIAL STEEL CO., LTD. , https://www.sdthreadedrods.com
(1) Choice of atmosphere and pressure
Atmosphere and pressure can affect the equilibrium temperature and peak shape of the sample chemical reaction and physical change. Therefore, the appropriate atmosphere and pressure must be selected according to the nature of the sample. Some samples are easily oxidized, and inert gases such as N2 and Ne can be introduced.
(2) Influence and Choice of Heating Rate
The heating rate not only affects the position of the peak temperature, but also affects the size of the peak area. In general, the peak area becomes larger and the peak becomes sharper at a faster heating rate. However, the rate of rapid heating increases the extent to which the sample decomposes away from the equilibrium conditions, making it easy to drift the baseline. More importantly, it may cause the adjacent two peaks to overlap and the resolution decreases. The slower rate of temperature rise, the smaller baseline drift, the closer the system to equilibrium conditions, the wider and shallower peaks, and the better separation of the adjacent two peaks, resulting in higher resolution.
(3) Pretreatment and dosage of the sample
Large amount of sample, easy to overlap the adjacent two peaks, reducing the resolution. Generally reduce the dosage as much as possible up to milligrams. The particle size of the sample is between 100 mesh and 200 mesh. Small particles can improve the heat conduction conditions, but too fine may destroy the crystallinity of the sample. For samples that are easily decomposed to generate gas, the particles should be larger. The granules, loading conditions and tightness of the reference should be consistent with the sample to reduce baseline drift.
(4) The choice of reference
For a smooth baseline, the choice of reference is important. It is required that the reference material does not undergo any change during heating or cooling. The specific heat, thermal conductivity, and particle size of the reference material should be the same as or similar to the sample during the entire temperature increase process.
(5) Selection of paper speed
Under the same experimental conditions, the same sample, such as the speed of paper feeding, the area of ​​the peak is large, but the shape of the peak is flat, the error is small; the speed of paper feeding is small and the area of ​​the peak is small. Therefore, according to different samples to choose the appropriate speed of paper. The choice of different conditions will affect the differential heat curve, in addition to the above there are many factors, such as the sample tube material, size and shape, the thermocouple material and the location of the thermocouple inserted in the sample and reference.
Differential thermal analysis often finds in the actual work that the same sample is measured on different instruments, or that different people measure on the same instrument. The resulting differential heat curve results are different. The main reason is that heat is related to many factors, and the heat transfer conditions are more complicated. It is generally caused by instruments and samples. The specific analysis of its factors is divided into five areas: