All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.
Â
Features for MicroPette - Lightweight, ergonomic, low force design - Digital display clearly reads volume setting - The pipettes cover volume range of 0.1μl to 10ml - Easy to calibrate and maintain with tool supplied - Design helps avoid repetitive strain injuries - Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate - The low part is available for autoclaving - Manufactured from innovative materials
Features for multi-channel MicroPette - 8 and 12 channel pipettes are available for standard 96-well plate - Dispensing head rotates for optimum pipetting convenience - Individual piston and tip cone assemblies allow easy repair and maintenance - Compound material tip cone design allows visual seal verification - Can be used with universal style pipette tips
Calibration
Please visit our website for online calibration at www.dragon-lab.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation. Â We will support and help you to achieve consistently excellent results.
Specifications
 Â
Single-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
0.1-2.5μL
0.05μL
2.5μL
2.50%
0.0625
2.00%
0.05
1.25μL
3.00%
0.0375
3.00%
0.0375
0.25μL
12.00%
0.03
6.00%
0.015
0.5-10μL
0.1μL
10μL
1.00%
0.1
0.80%
0.08
5μL
1.50%
0.075
1.50%
0.075
1μL
2.50%
0.025
1.50%
0.015
2-20μL
0.5μL
20μL
0.90%
0.18
0.40%
0.08
10μL
1.20%
0.12
1.00%
0.1
2μL
3.00%
0.06
2.00%
0.04
5-50μL
0.5μL
50μL
0.60%
0.3
0.30%
0.15
25μL
0.90%
0.225
0.60%
0.15
5μL
2.00%
0.1
2.00%
0.1
10-100μL
1μL
100μL
0.80%
0.8
0.15%
0.15
50μL
1.00%
0.5
0.40%
0.2
10μL
3.00%
0.3
1.50%
0.15
20-200μL
1μL
200μL
0.60%
1.2
0.15%
0.3
100μL
0.80%
0.8
0.30%
0.3
20μL
3.00%
0.6
1.00%
0.2
50-200μL
1μL
200μL
0.60%
1.2
0.15%
0.3
100μL
0.80%
0.8
0.30%
0.3
50μL
1.00%
0.5
0.40%
0.2
100-1000μL
5μL
1000μL
0.60%
6
0.20%
2
500μL
0.70%
3.5
0.25%
1.25
100μL
2.00%
2
0.70%
0.7
200-1000μL
5μL
1000μL
0.60%
6
0.20%
2
500μL
0.70%
3.5
0.25%
1.25
200μL
0.90%
1.8
0.30%
0.6
1000-5000μL
50μL
5000μL
0.50%
25
0.15%
7.5
2500μL
0.60%
15
0.30%
7.5
1000μL
0.70%
7
0.30%
3
2-10mL
0.1mL
10mL
0.60%
60
0.20%
20
5mL
1.20%
60
0.30%
15
2mL
3.00%
60
0.60%
12
8-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
0.5-10μL
0.1μL
10μL
1.50%
0.15
1.50%
0.15
5μL
2.50%
0.125
2.50%
0.125
1μL
4.00%
0.04
4.00%
0.04
5-50μL
0.5μL
50μL
1.00%
0.5
0.50%
0.25
25μL
1.50%
0.375
1.00%
0.25
5μL
3.00%
0.15
2.00%
0.1
50-300μL
5μL
300μL
0.70%
2.1
0.25%
0.75
150μL
1.00%
1.5
0.50%
0.75
50μL
1.50%
0.75
0.80%
0.4
12-channel Adjustable Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
0.5-10μL
0.1μL
10μL
1.50%
0.15
1.50%
0.15
5μL
2.50%
0.125
2.50%
0.125
1μL
4.00%
0.04
4.00%
0.04
5-50μL
0.5μL
50μL
1.00%
0.5
0.50%
0.25
25μL
1.50%
0.375
1.00%
0.25
5μL
3.00%
0.15
2.00%
0.1
50-300μL
5μL
300μL
0.70%
2.1
0.25%
0.75
150μL
1.00%
1.5
0.50%
0.75
50μL
1.50%
0.75
0.80%
0.4
Fixed Volume Pipettes
Volume Range
Increment
Test Volume
Maximum permissible systematic error (Inaccuracy)
Maximum permissible random error (Imprecision)
%
μL
%
μL
5μL
-
5μL
1.3%
0.065
1.2%
0.06
10μL
-
10μL
0.8%
0.08
0.8%
0.08
20μL
-
20μL
0.6%
0.12
0.5%
0.1
25μL
-
25μL
0.5%
0.125
0.3%
0.075
50μL
-
50μL
0.5%
0.25
0.3%
0.15
100μL
-
100μL
0.5%
0.5
0.3%
0.3
200μL
-
200μL
0.4%
0.8
0.2%
0.4
250μL
--
250μL
0.4%
1.0
0.2%
0.5
500μL
-
500μL
0.3%
1.5
0.2%
1.0
1000μL
-
1000μL
0.3%
3.0
0.2%
2.0
2000μL
-
2000μL
0.3%
6.0
0.15%
3.0
5000μL
-
5000μL
0.3%
15
0.15%
7.5
    Â
Â
Mechanical Pipettes (Adjustable & Fixed)
All pipettes have been quality tested according to ISO8655-2:2002 with calibration certificate. The quality control involves gravimetric testing of each pipette with distilled water at 22ºC.
Â
Features for MicroPette - Lightweight, ergonomic, low force design - Digital display clearly reads volume setting - The pipettes cover volume range of 0.1μl to 10ml - Easy to calibrate and maintain with tool supplied - Design helps avoid repetitive strain injuries - Calibrated in accordance with ISO8655. Each pipette supplied with individual test certificate - The low part is available for autoclaving - Manufactured from innovative materials
Features for multi-channel MicroPette - 8 and 12 channel pipettes are available for standard 96-well plate - Dispensing head rotates for optimum pipetting convenience - Individual piston and tip cone assemblies allow easy repair and maintenance - Compound material tip cone design allows visual seal verification - Can be used with universal style pipette tips
Calibration
Please visit our website for online calibration at www.dragon-lab.com. Through the online calibration software, users can perform simple, accurate and timely calibration with free cost, and avoid the calibration error due to non-professional operation. Â We will support and help you to achieve consistently excellent results.